Kinetic analysis of IgG antibodies to beta-amyloid oligomers with surface plasmon resonance

Publication Date

7-15-2015

Document Type

Article

Abstract

Surface plasmon resonance was used to investigate the kinetics, affinity, and specificity of binding between anti-Aβ (beta-amyloid) IgG antibodies and oligomeric Aβ. Two factors were needed to accurately characterize the IgG binding kinetics. First, a bivalent model was necessary to properly fit the kinetic association and dissociation sensograms. Second, a high concentration of IgG was necessary to overcome a significant mass transport limitation that existed regardless of oligomer density on the sensor surface. Using high IgG concentrations and bivalent fits, consistent kinetic parameters were found at varying sensor surface ligand densities. A comparison of binding specificity, affinity, and kinetic flux between monoclonal and natural human anti-Aβ IgG antibodies revealed the following findings. First, monoclonal antibodies 6E10 and 4G8 single-site binding affinity is similar between Aβ oligomers and monomers. Second, natural human anti-Aβ IgG binding readily binds Aβ oligomers but does not bind monomers. Third, natural human anti-Aβ IgG binds Aβ oligomers with a higher affinity and kinetic flux than 6E10 and 4G8. Both the current analytical methodology and antibody binding profiles are important for advances in antibody drug development and kinetic biomarker applications for Alzheimer’s disease.

Publication Title

Analytical Biochemistry

Volume

481

First Page

43

Last Page

54

DOI

10.1016/j.ab.2015.03.032

Publisher Policy

pre-print, post-print

This document is currently not available here.

Find in your library

Share

COinS